Non-invasive measurement and validation of tissue oxygen saturation covered with overlying tissues

  • 期刊名字:自然科学进展(英文版)
  • 文件大小:
  • 论文作者:Yichao Teng,Haishu Ding,Lan Hu
  • 作者单位:Department ofBiomedical Engineering,China,Department of Clinical Laboratories,Graduate School at Shenzhen,Department of
  • 更新时间:2022-11-24
  • 下载次数:
论文简介

In this paper,the biological tissue oxygen saturation(rS02)is obtained non-invasively and in real time based on near infrared spec-troscopy(NIRS)using two emitting wavelengths and two detectors,where the tissue is covered with overlying tissues.Our group devel-oped an NIRS oximeter based on the above principle independently,and validated it using liquid tissue model calibrations and animal experiments.The results indicate that(1)in the normal range of tissue oxygen saturation(40-70%),the rS02 measured by NIRS is accu-rate enough and little influenced by the background absorptions(such as the absorption of water)and overlying tissues(such as fat);(2)during cerebral hypoxia and recovery of three piglets,there is excellent correlation(p