Experimental study of multi-photon contamination on the measurement of fluorescent decay time Experimental study of multi-photon contamination on the measurement of fluorescent decay time

Experimental study of multi-photon contamination on the measurement of fluorescent decay time

  • 期刊名字:中国物理C(英文版)
  • 文件大小:
  • 论文作者:ZHU Wen-Han,ZHAO Yan-E,ZHANG Y
  • 作者单位:Department of Modern Physics
  • 更新时间:2022-11-24
  • 下载次数:
论文简介

In the measurement of fluorescent lifetime based on time correlation-single photon counting technique by means of TAC, due to the contamination of multi-photons a deviation of fluorescent lifetime measured from the expected value is experimentally studied. A correction function instead of a simple exponential function is used to fit the experiment data. The validation of the correction function is checked using the experimental data of several test samples: YAP, NaI(T1) and LSO. The results show that the correction function well fits the data and the reasonable fluorescent lifetimes are obtained.

论文截图
版权:如无特殊注明,文章转载自网络,侵权请联系cnmhg168#163.com删除!文件均为网友上传,仅供研究和学习使用,务必24小时内删除。