						Evaluation of Atypical Lymphocyte Warnings of Sysmex XE-4000 in Outpatient Children
- 期刊名字:农业科学与技术(英文版)
- 文件大小:
- 论文作者:Xueling GUO,Tingbo ZHANG,Maofa
- 作者单位:Department of Examination,Department of Clinical Laboratory
- 更新时间:2023-04-17
- 下载次数:次
This study aimed to evaluate the atypical lymphocyte warnings of Sysmex XE-4000 in outpatient children. [Method] Randomly 280 specimens of outpatient children were selected to analyze atypical lymphocytes with Sysmex XE-4000 and artificial smear microscopy simultaneously. [Result] With artificial smear microscopy as the gold standard, the atypical lymphocyte warnings of Sysmex XE-4000 in outpatient children exhibited a sensitivity of 97.4%, specificity of 69.3%, positive predictive value (PPV) of 55%, and negative predictive value (NPV) of 98.5%. [Conclusion] Sysmex XE-4000 shows a high sensitivity in atypical lymphocyte warnings, which can reduce the number of blood smears, improve the specificity of microscopy, and decline the labor intensity of inspection personnel. However, there are stil certain false positives to be confirmed by artificial microscopy.
- 