Measurement of boron isotopes by negative thermal ionization mass spectrometry Measurement of boron isotopes by negative thermal ionization mass spectrometry

Measurement of boron isotopes by negative thermal ionization mass spectrometry

  • 期刊名字:核技术
  • 文件大小:
  • 论文作者:LIU Wei-Guo,PENG Zi-cheng,XIAO
  • 作者单位:Department of Earth and Space Sciences,Qinghai Institute of Salt Lakes
  • 更新时间:2022-11-24
  • 下载次数:
论文简介

The isobaric interference for boron isotopic measurement by negative thermal ionization mass spectrometry (NTIMS) has been studied. The result shows that the CNO- is not only from the organic material, but also from nitrate in loading reagent in NTIMS. Monitoring the mass 43 ion intensity and 43/42 ratio of blank are also necessary for the boron isotopic measurement by NTIMS, other than is only boron content.

论文截图