In-situ measurement of the dissolved S2- in seafloor diffuse flow system: sensor preparation and cal In-situ measurement of the dissolved S2- in seafloor diffuse flow system: sensor preparation and cal

In-situ measurement of the dissolved S2- in seafloor diffuse flow system: sensor preparation and cal

  • 期刊名字:浙大学报(英文版)(A辑:应用物理和工程)
  • 文件大小:
  • 论文作者:Ying YE,Xia HUANG,Yi-wen PAN,C
  • 作者单位:Department of Earth Sciences,The State Key Laboratory of Fluid Power Transmission & Control
  • 更新时间:2022-11-24
  • 下载次数:
论文简介

The preparation approach and calibration result of an improved type of ion selective electrode (ISE), which is used to measure the total dissolved S2-, are introduced in this paper. The improved Ag/Ag2S electrode uses silver wire as the substrate, which is surrounded by electric polymer containing superfine silver powder. After the stabilization of the epoxy-resin, Ag2ES layer was formed by chemical reaction with 0.2 mol/L (NH4)2S solution for 5 min. With Ag/AgCl as reference electrode, the Ag/Ag2S electrode can be used to measure dissolved S2-. The correlation between the measured potentials and the logarithm of dissolved S2- is found to be linear, within range of the concentration of dissolved S2- from 10-2~10-7 mol/L. The slope of the regression line between measured potential and logarithm of dissolved S2- is about -27.7, which agrees well with the theoretical Nernst value -29.6. Furthermore, the performance of the improved Ag/Ag2S electrode, such as the response time, sensitivity and stability, greatly outweighs the conventional Ag/Ag2S electrode.