On-line parallel reversed phase two-dimensional liquid chromatography for high-throughput analysis o On-line parallel reversed phase two-dimensional liquid chromatography for high-throughput analysis o

On-line parallel reversed phase two-dimensional liquid chromatography for high-throughput analysis o

  • 期刊名字:中国科学B辑(英文版)
  • 文件大小:
  • 论文作者:WANG Zhicong,ZHANG Qinghe,LI T
  • 作者单位:Dalian Institute of Chemical Physics,Faculty of Material Science and Chemical Engineering,Dalian Elite Analytical Instru
  • 更新时间:2023-01-17
  • 下载次数:
论文简介

A comprehensive two-dimensional liquid chromatographic system (2D SCX/RP) is constructed with a 10-port-2-way valve using strong cation exchange chromatography (Hypersil SCX, 100 mm×4.6 mm I.D.) followed by reversed phase chromatography (Hypersil BDS C18, 15 mm×4.6 mm I.D.) to separate the complex peptides from globin peptic hydrolysate. After the sample was loaded on the SCX column, the phosphate buffer (pH 4.0) was used to elute the peptides. Then, elutes flowed through the interface and the peptides focused on the head of the trapping columns (Hypersil BDS C18, 15 mm×4.6 mm I.D.) but salt passed into the waste. After the valve was switched, the samples were flushed with a backward flow into the RP analytical column. The peptides on the SCX were eluted with 12 discontinuous steps linearly increasing salt concentrations. The peptides enriched on the trapping column were desalted and separated by the RP columns. The resolution and the resolved peaks of the 2D SCX/RP system were greatly increased and the total peak capacity reached as high as 2280.

论文截图