Design and testing of micro fluidic chemical analysis chip integrated with micro valveless pump
- 期刊名字:中国科学E辑(英文版)
- 文件大小:
- 论文作者:FU Xin,XIE Haibo,Yang Huayong,
- 作者单位:State Key Laboratory of Fluid Power Transmission and Control,Institute of Microanalytical Systems
- 更新时间:2022-12-26
- 下载次数:次
A new structure and working principle of the chip integrated with micro valveless pump for capillary electrophoresis was proposed in this paper. The micro valveless pump with plane structure has advantages of simple structure, and the process technology is compatible with existing micro chips for capillary electrophoresis. Based upon the mathematical model, simulation study of micro pump was carried out to investigate the influence of structural parameters on flow characteristics, and the performance of the integrated micro pump was also tested with different control parameters. The simulation results agree with the experimental results. Three samples, which are amino acid, fluorescein and buffer solution, have been examined with this chip. The results of the primary experiments showed that the micro valveless pump was promising in the integration and automatization of miniature integrated fluidic systems.
-