首页 > 标准下载>IEC 61010-2-101-2015 测量、控制和实验室用电气设备的安全要求--第2-101部分:体外诊断(IVD)医疗设备的特殊要求 Safety requirements for electrical equipment for measurement, control and laboratory use – Part 2-101: Particular requirements for in vitro diagnostic (IVD) medical equipment免费下载
IEC 61010-2-101-2015 测量、控制和实验室用电气设备的安全要求--第2-101部分:体外诊断(IVD)医疗设备的特殊要求 Safety requirements for electrical equipment for measurement, control and laboratory use – Part 2-101: Particular requirements for in vitro diagnostic (IVD) medical equipment IEC 61010-2-101-2015 测量、控制和实验室用电气设备的安全要求--第2-101部分:体外诊断(IVD)医疗设备的特殊要求 Safety requirements for electrical equipment for measurement, control and laboratory use – Part 2-101: Particular requirements for in vitro diagnostic (IVD) medical equipment

IEC 61010-2-101-2015 测量、控制和实验室用电气设备的安全要求--第2-101部分:体外诊断(IVD)医疗设备的特殊要求 Safety requirements for electrical equipment for measurement, control and laboratory use – Part 2-101: Particular requirements for in vitro diagnostic (IVD) medical equipment

  • 标准类别:
  • 标准大小:
  • 标准编号:IEC 61010-2-101-2015
  • 标准状态:现行
  • 更新时间:2023-08-28
  • 下载次数:
标准简介

Replace the text by the following:
This part of IEC 61010 applies to equipment intended for in vitro diagnostic (IVD) medical
purposes, including self-test IVD medical purposes.
IVD medical equipment, whether used alone or in combination, is intended by the
manufacturer to be used in vitro for the examination of specimens, including blood and tissue
samples, derived from the human body, solely or principally for the purpose of providing
information concerning one or more of the following:
• a physiological or pathological state; or
• a congenital abnormality;
• the determination of safety and compatibility with potential recipients;
• the monitoring of therapeutic measures.
Self-test IVD medical equipment is intended by the manufacturer for use by lay persons in a
home environment.
NOTE If all or part of the equipment falls within the scope of one or more other part 2 standards of IEC 61010 as
well as within the scope of this standard, considerations have to be given to  those other part 2 standards.
1.1.2 Equipment excluded from scope
Addition:
Add the following item:
aa)  Equipment in the scope of IEC 61010-2-081 unless they are specifically intended by
their manufacturer to be used for in vitro diagnostic examination.dition 2.0 2015-01 INTERNATIONAL STANDARD NORME INTERNATIONALE GROUP SAFETY PUBLICATION PUBLICATION GROUPéE DE SéCURITé Safety requirements for electrical equipment for measurement, control and laboratory use – Part 2-101: Particular requirements for in vitro diagnostic (IVD) medical equipment Règles de sécurité pour appareils électriques de mesurage, de régulation et de laboratoire – Partie 2-101: Exigences particulières pour les appareils médicaux de diagnostic in vitro (DIV)